MS pipette while providing customers with the “Best Value” combination of quality, features, service and price.
Best Accuracy & Precision
By ISO calibration and certification, MS pipette offer accurate & precise results for your research.
Applicable to universal tip
MS pipette has compatibility with a majority of brands of tips on the market.
Comfort use experisence,
Your comfort through perfect fit, lightweight & ergonomic for use MS pipette, and very easy for single-hand operation.
Precise volume adjustment & light pipetting forces
Use the serial number to identify the pipette and to determine its age
* Suitable for major science Tip, MS-PW Series.

| Model | Volume range (µL) | Volume (µL) | Systematic error (µL) | Systematic error (%) | Random error (µL) | Random error (%) |
|---|---|---|---|---|---|---|
| MSP-2 | 0.2 - 2 | 0.2 | ± 0.026 | ± 13 | ≤ 0.013 | ≤ 6.5 |
| 2 | ± 0.033 | ± 1.65 | ≤ 0.015 | ≤ 0.75 | ||
| MSP-10 | 1 - 10 | 1 | ± 0.035 | ± 3.5 | ≤ 0.013 | ≤ 1.3 |
| 10 | ± 0.110 | ± 1.1 | ≤ 0.044 | ≤ 0.44 | ||
| MSP-20 | 2 - 20 | 2 | ± 0.11 | ± 5.5 | ≤ 0.033 | ≤ 1.65 |
| 20 | ± 0.20 | ± 1.0 | ≤ 0.066 | ≤ 0.33 | ||
| MSP-100 | 10 - 100 | 10 | ± 0.39 | ± 3.9 | ≤ 0.11 | ≤ 1.1 |
| 100 | ± 0.80 | ± 0.8 | ≤ 0.17 | ≤ 0.17 | ||
| MSP-200 | 20 - 200 | 20 | ± 0.55 | ± 2.75 | ≤ 0.22 | ≤ 1.1 |
| 200 | ± 1.60 | ± 0.8 | ≤ 0.33 | ≤ 0.17 | ||
| MSP-1000 | 100 - 1 000 | 100 | ± 3.3 | ± 3.3 | ≤ 0.7 | ≤ 0.7 |
| 1000 | ± 8.0 | ± 0.8 | ≤ 1.7 | ≤ 0.17 | ||
| MSP-5000 | 500 - 5 000 | 500 | ± 13 | ± 2.6 | ≤ 3.3 | ≤ 0.66 |
| 5000 | ± 33 | ± 0.66 | ≤ 8.8 | ≤ 0.18 | ||
| MSP-10000 | 1000 - 10 000 | 1000 | ± 33 | ± 3.3 | ≤ 6.6 | ≤ 0.66 |
| 10000 | ± 60 | ± 0.6 | ≤ 17.6 | ≤ 0.18 |
| Model | Volume range (µl) | Volume (µl) | Systematic error (µl) | Systematic error (%) | Random error (µl) | Random error (%) |
|---|---|---|---|---|---|---|
| MSP-8X10 / MSP-12X10 | 0.5-10 | 0.5 | ± 0.09 | ± 18 | ≤ 0.04 | ≤ 8.8 |
| 10 | ± 0.22 | ± 2.2 | ≤ 0.11 | ≤ 1.1 | ||
| MSP-8X20 / MSP-12X20 | 2-20 | 2 | ± 0.11 | ± 5.5 | ≤ 0.09 | ≤ 4.5 |
| 20 | ± 0.40 | ± 2.0 | ≤ 0.17 | ≤ 0.85 | ||
| MSP-8X200 / MSP-12X200 | 20-200 | 20 | ± 0.55 | ± 2.75 | ≤ 0.28 | ≤ 1.4 |
| 200 | ± 2.20 | ± 1.1 | ≤ 0.55 | ≤ 0.275 | ||
| MSP-8X300 / MSP-12X300 | 20-300 | 20 | ± 1.10 | ± 5.5 | ≤ 0.39 | ≤ 1.95 |
| 300 | ± 3.30 | ± 1.1 | ≤ 1.10 | ≤ 0.37 |
| Tip Compatibility | ||||||||
|---|---|---|---|---|---|---|---|---|
| major science Single Channel Pipettes w/ Tips | ||||||||
| Key: | Ο:Not Compatible Ο:Poor Fit Ο:Ok fit Ο:Good fit -: N/A Φ: No ejection possible | |||||||
| major science Pipettes | ||||||||
| Tip Brand | Volume | MSP-2, MSP-10 | MSP-20 | MSP-100 | MSP-200 | MSP-1000 | MSP-5000 | MSP-10000 |
| Axygen | 10µL (short) | Ο | Φ | - | - | - | - | - |
| 10µL (long) | Ο | Φ | - | - | - | - | - | |
| 20µL | Ο | Ο | - | - | - | - | - | |
| 100µL | - | - | Ο | Ο | - | - | - | |
| 200µL | - | Ο | Ο | Ο | - | - | - | |
| 300µL | - | - | - | Ο | - | - | - | |
| 1000µL | - | - | - | - | Ο | - | - | |
| 1100µL | - | - | - | - | Ο | - | - | |
| 5000µl | - | - | - | - | - | Ο | - | |
| Brand | 10µl (long) | Ο | Φ | - | - | - | - | - |
| 200µl | - | - | - | - | ||||
| 1000µl | - | - | - | - | Ο | - | - | |
| 5000µl | - | - | - | - | - | Ο | - | |
| 10mL | - | - | - | - | - | - | Ο | |
| Greiner / Sapphire | 10µL | Ο | Φ | - | - | - | - | - |
| 20µL | Ο | Φ | - | - | - | - | - | |
| 100µL | - | - | Ο | - | - | - | - | |
| 200µL | - | - | Ο | Ο | - | - | - | |
| 1000µL | - | - | - | - | Ο | - | - | |
| Starlab | 10µL (long) | Ο | Φ | - | - | - | - | - |
| 20µL | Ο | Ο | Ο | - | - | - | - | |
| 100µL | - | - | Ο | - | - | - | - | |
| 200µL | - | Ο | Ο | Ο | - | - | - | |
| 300µL | - | - | Ο | Ο | - | - | - | |
| 1000µL | - | - | - | - | Ο | - | - | |
| 1250µl | - | - | - | - | - | - | - | |
| 5000µl | - | - | - | - | - | Ο | - | |
| 10mL | - | - | - | - | Ο | - | Ο | |
| Eppendorf | 10µL (long) | Ο | Φ | - | - | - | - | - |
| 20µL | Ο | Ο | Ο | - | - | - | - | |
| 100µL | - | - | Ο | Ο | - | - | - | |
| 200µL | - | Ο | Ο | Ο | - | - | - | |
| 300µL | - | - | - | Ο | - | - | - | |
| 1000µL | - | - | - | - | Ο | - | - | |
| 1250µl | - | - | - | - | - | - | - | |
| 5000µl | - | - | - | - | - | Ο | - | |
| 10mL | - | - | - | - | - | - | Ο | |
| Sartorius | 10µL | Ο | Φ | - | - | - | - | - |
| 200µl | - | Ο | Ο | Ο | - | - | - | |
| 350µL | - | - | Ο | Ο | - | - | - | |
| 1000µl | - | - | - | - | Ο | - | - | |
| 5000µl | - | - | - | - | - | Ο | - | |
| 10mL | - | - | - | - | - | - | Ο | |
| Rainin | 10µL (short) | Ο | Φ | - | - | - | - | - |
| 10µL (long) | Ο | Φ | - | - | - | - | - | |
| 20µL | Ο | Ο | - | - | - | - | - | |
| 100µL | - | - | Ο | Ο | - | - | - | |
| 200µL | - | Ο | - | Ο | - | - | - | |
| 350µL | - | - | Ο | Ο | - | - | - | |
| 1000µl | - | - | - | - | Ο | - | - | |
| 5000µl | - | - | - | - | - | Ο | - | |
| 10mL | - | - | - | - | - | - | Ο | |
| Sorenson Expert | 10µL (short) | Ο | Φ | - | - | - | - | - |
| 10µL (long) | Ο | Φ | - | - | - | - | - | |
| 200µL | - | Ο | Ο | Ο | - | - | - | |
| 300µL | - | - | - | Ο | - | - | - | |
| 1250µl | - | - | - | - | Ο | - | - | |
| Labcon | 10µL | Ο | Φ | - | - | - | - | - |
| 200µL | - | - | Ο | Ο | - | - | - | |
| 300µL | - | - | - | Ο | - | - | - | |
| 1000µl | - | - | - | - | Ο | - | - | |
| Neptune | 10µL | Ο | Φ | - | - | - | - | - |
| 20µl | Ο | Φ | - | - | - | - | - | |
| 100µl | - | - | Ο | - | - | - | - | |
| 200µl | - | - | Ο | Ο | - | - | - | |
| 300µl | - | - | - | Ο | - | - | - | |
| 1000µl | - | - | - | - | Ο | - | - | |
| 5000µl | - | - | - | - | - | Ο | - | |
| Biotix | 10µl | Ο | Φ | - | - | - | - | - |
| 20µl | Ο | Ο | - | - | - | - | - | |
| 100µl | - | - | Ο | - | - | - | - | |
| 200µl | - | - | Ο | Ο | - | - | - | |
| 300µl | - | - | - | Ο | - | - | - | |
| 1250µl | - | - | - | - | Ο | - | - | |
| Thermo Art | 10µl | Ο | - | - | - | - | - | - |
| 200µl | - | Ο | Ο | Ο | - | - | - | |
| 300µl | - | Ο | - | Ο | - | - | - | |
| 1000µl | - | - | - | - | Ο | - | - | |
| 1250µl | - | - | - | - | Ο | - | - | |
| Thermo Finntip | 10µl (short) | Ο | - | - | - | - | - | - |
| 10µl (long) | - | Ο | - | - | - | - | - | |
| 200µl | - | Ο | Ο | Ο | - | - | - | |
| 300µl | - | - | - | Ο | - | - | - | |
| 1000µl | - | - | - | - | Ο | - | - | |
| 5000µl | - | - | - | - | - | Ο | - | |
| 10mL | - | - | - | - | - | - | Ο | |
| Cat. No | Description |
|---|---|
| MSP-2 | MS pipette, variable volume 0.2~2μl |
| MSP-10 | MS pipette, variable volume 1~10μl |
| MSP-20 | MS pipette, variable volume 2~20μl |
| MSP-100 | MS pipette, variable volume 10~100μl |
| MSP-200 | MS pipette, variable volume 20~200μl |
| MSP-1000 | MS pipette, variable volume 100~1000μl |
| MSP-5000 | MS pipette, variable volume 500~5000μl |
| MSP-10000 | MS pipette, variable volume 1000~10000μl |
| MSP-8X10 | MS 8-ch pipette, variable volume 0.5~10μl |
| MSP-8X20 | MS 8-ch pipette, variable volume 2~20μl |
| MSP-8X200 | MS 8-ch pipette, variable volume 20~200μl |
| MSP-8X300 | MS 8-ch pipette, variable volume 20~300μl |
| MSP-12X10 | MS 12-ch pipette, variable volume 0.5~10μl |
| MSP-12X20 | MS 12-ch pipette, variable volume 2~20μl |
| MSP-12X200 | MS 12-ch pipette, variable volume 20~200μl |
| MSP-12X300 | MS 12-ch pipette, variable volume 20~300μl |
| MSP-ST03 | Pipette Stand |
*Technical specification and price not final and are subject to change.
MSP-ST02
| Cat No. | Volume (µL) | SPECIFICATIONS | ISO 8655 | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Systematic error | Random error | Systematic error | Random error | ||||||
| (µL) | (%) | (µL) | (%) | (µL) | (%) | (µL) | (%) | ||
| MSP-2 | 0.2 | ±0.026 | ±13.2% | ≤0.013 | ≤6.6% | ±0.080 | ±40.0% | ≤0.040 | ≤20.0% |
| 2 | ±0.033 | ±1.7% | ≤0.015 | ≤0.8% | ±0.080 | ±4.0% | ≤0.040 | ≤2.0% | |
| MSP-10 | 1 | ±0.028 | ±2.8% | ≤0.013 | ≤1.3% | ±0.120 | ±12.0% | ≤0.080 | ≤0.8% |
| 10 | ±0.110 | ±1.1% | ≤0.044 | ≤0.4% | ±0.120 | ±1.2% | ≤0.080 | ≤0.8% | |
| MSP-20 | 2 | ±0.11 | ±5.5% | ≤0.033 | ≤1.7% | ±0.20 | ±10.0% | ≤0.100 | ≤5.0% |
| 20 | ±0.20 | ±1.0% | ≤0.066 | ≤0.3% | ±0.20 | ±1.0% | ≤0.100 | ≤0.5% | |
| MSP-100 | 10 | ±0.39 | ±3.9% | ≤0.11 | ≤1.1% | ±0.80 | ±8.0% | ≤0.030 | ≤3.0% |
| 100 | ±0.80 | ±0.8% | ≤0.17 | ≤0.2% | ±0.80 | ±8.0% | ≤0.030 | ≤0.3% | |
| MSP-200 | 20 | ±0.55 | ±2.8% | ≤0.22 | ≤1.1% | ±1.60 | ±8.0% | ≤0.60 | ≤3.0% |
| 200 | ±1.60 | ±0.8% | ≤0.33 | ≤0.2% | ±1.60 | ±0.8% | ≤0.60 | ≤0.3% | |
| MSP-1000 | 100 | ±3.3 | ±3.3% | ≤0.7 | ≤0.7% | ±8.0 | ±8.0% | ≤3.0 | ≤3.0% |
| 1000 | ±0.8 | ±8.0% | ≤1.7 | ≤0.2% | ±8.0 | ±0.8% | ≤3.0 | ≤0.3% | |
| MSP-5000 | 500 | ±13 | ±2.6% | ≤3.3 | ≤0.7% | ±40 | ±8.0% | ≤15 | ≤3.0% |
| 5000 | ±33 | ±0.7% | ≤8.8 | ≤0.2% | ±40 | ±0.8% | ≤15 | ≤0.3% | |
| MSP-10000 | 1000 | ±33 | ±3.3% | ≤6.6 | ≤0.7% | ±60 | ±6.0% | ≤30 | ≤3.0% |
| 10000 | ±60 | ±0.6% | ≤17.6 | ≤0.2% | ±60 | ±0.6% | ≤30 | ≤0.3% | |
| MSP-8X10 | 0.5 | ±0.09 | ±17.6% | ≤0.04 | ≤8.8% | ±0.24 | ±48.0% | ≤0.16 | ≤32.0% |
| MSP-8X20 | 2 | ±0.11 | ±5.5% | ≤0.09 | ≤4.4% | ±0.40 | ±20.0% | ≤0.20 | ≤10.0% |
| MSP-8X200 | 20 | ±0.6 | ±2.8% | ≤0.28 | ≤1.4% | ±3.2 | ±16.0% | ≤1.20 | ≤6.0% |
| MSP-8X300 | 20 | ±1.1 | ±3.7% | ≤0.39 | ≤1.3% | ±8.0 | ±26.7% | ≤3.00 | ≤10.0% |
| MSP-12X10 | 10 | ±0.22 | ±2.2% | ≤0.11 | ≤1.1% | ±0.24 | ±2.4% | ≤0.16 | ≤1.6% |
| MSP-12X20 | 20 | ±0.40 | ±2.0% | ≤0.17 | ≤0.8% | ±0.40 | ±2.0% | ≤0.20 | ≤1.0% |
| MSP-12X200 | 200 | ±2.2 | ±1.1% | ≤0.55 | ≤0.3% | ±3.2 | ±1.6% | ≤1.20 | ≤0.6% |
| MSP-12X300 | 300 | ±3.3 | ±1.1% | ≤1.10 | ≤0.4% | ±8.0 | ±2.7% | ≤3.00 | ≤1.0% |
![]() |
![]() |
![]() |
![]() |
| MSP-2 | MSP-200 | MSP-8X10 | MSP-12X10 |
![]() |
![]() |
![]() |
![]() |
| MSP-100 | MSP-1000 | MSP-8X20 | MSP-12X20 |
![]() |
![]() |
![]() |
![]() |
| MSP-20 | MSP-5000 | MSP-8X200 | MSP-12X200 |
![]() |
![]() |
![]() |
![]() |
| MSP-100 | MSP-10000 | MSP-8X300 | MSP-12X300 |